Microscopic examination is the final verification step of specimen preparation. The objective parameters and observation modes of a metallurgical microscope determine which microstructural details can be resolved; proper selection provides reliable evidence for interpretation without sacrificing efficiency.
Objective Parameters and Selection
The core parameters of an objective are magnification and numerical aperture (NA). Resolution is approximately 0.61λ/NA, and any magnification beyond roughly 1000×NA is empty magnification. Objectives should be configured on the principle of low-power survey followed by high-power inspection: 5× and 10× for macroscopic surveys of structure and defects, 20× and 50× for routine microstructural rating, and a 100× oil-immersion objective only when fine precipitates must be resolved. Plan-achromat objectives meet most inspection needs; plan-apochromat objectives are preferred where contrast and chromatic correction are critical.
Brightfield and Darkfield
Brightfield (BF) is the basic mode: light passes through the objective onto the specimen at normal incidence, so flat reflective regions appear bright and scattering regions dark. It suits the vast majority of etched structures for examination and rating. Darkfield (DF) collects only the light scattered by the specimen, rendering scratches, inclusions, and fine relief as bright features. It is sensitive to polishing defects and shallow relief, and is often used to check preparation quality and non-metallic inclusions.
Polarized Light and DIC
Polarized light (POL) uses crossed polarizers to examine anisotropic materials; grain-orientation contrast is pronounced in anodized aluminum coatings, titanium, zinc, and magnesium, while for isotropic metals it can reveal stress and deformation traces. Differential interference contrast (DIC) converts minute surface height differences into contrast, producing a relief-like three-dimensional image; it suits unetched or lightly etched specimens for observing slip bands, surface relief, and hardness indentations. Note that DIC demands a highly flat specimen, so preparation quality directly affects image quality.
Practical Tips
- Configure objectives for low-power survey and high-power inspection; avoid empty magnification.
- Use brightfield for routine rating and darkfield for inclusion and scratch inspection.
- Polarized light suits anisotropic materials; DIC suits relief observation on unetched surfaces.
- Confirm flatness and adequate etching before examination; specimen evaluation and the actual operating procedure shall prevail.